In this lab you will be mapping the electric field of the EOD in a given
volume of water. Before beginning the setup, the actual area covered
must be determined. We
measured a section of tank 14 cm in length and 14 in width, which easily
sufficed our
needs. We also suspended the fish approximately 15 cm deep in the water.
Materials:
Twenty thin wooden skewers
Needle
Thread
String
Permanent marker
Masking tape
Pantyhose: unlaundered with a fairly loose weave
Electrical wire: shielded, coaxial, and insulated
Solid glass rod with diameter between 1 and 3 cm
Live weakly electric fish
Solid metal rod with diameter between 1 and 3 cm
55 gallon tank
Plexiglass: fitting within the tank and at least the size of the area
to be mapped
Amplifier
preamplifier
Electrode
Oscilliscope
Measurement System:
Anchor point
On one lengthwise side of tank, choose a point
(0,0) in the tank at which the center of the fish
will
be. The fish will be located at the center of the
width. You will use this Anchor point to
correctly position the grid, the X, and Y markers
and
the suspension apparatus.
Plexiglas grid:
On a piece of Plexiglas, mark off an x and
y-axis so that (0,0) is at the center of the glass
with a wax pencil or permanent marker.
Mark off lines every 2 cm in each direction as
far as will be measured.
Place the grid on the bottom of the tank.
For the x coordinates:
On masking tape, mark off a center point, and
then mark off 2 cm increments in the positive and
negative direction for the length that will be
measured (the length of the grid).
Place the strip of tape along the length
of the tank, SO THAT X=0 CORRESPONDS WITH YOUR
ANCHOR
POINT (0,0).
Put an identical piece of tape on each side
where measurements will be taken.
For the y coordinates:
On masking tape, mark off a center point and
2 cm increments in the positive and negative
direction
for the length that will be measured.
Place each end of the tape on one side of the
tank
so that the tape extends across the width of the
tank.
Make sure that the tape is slightly outside of
where
all measurements will be taken. Also,
the center point of the tape should be directly
above the
x axis of the plexiglass grid.
Put an identical piece of tape on each side
where measurements will be taken.
String
Use two pieces of string, slightly longer
than the length and width of the area being
measured.
The strings will be used to mark the point being
measured (description in procedure)
Electrode Setup
Creating the electrode
Wind two insulated wires and run them down the
length of a thin wooden skewer.
Strip the end of each wire that will be in the
water (only strip a small portion of each wire).
Make
sure stripped portions begin beyond the end of the
skewer.
Separate the stripped ends of the wires so
that they are about 1 cm apart.
Mark the skewer at the point that is level
with the water line when the stripped points of
the
electrode are level with the mid-line of the
fish.
Connection to coaxial wire
Strip the ends of the wires that are not going
into the water.
Solder one of the stripped ends to the ground
(outer) wire within the coaxial wire (have your
teacher
demonstrate). This wire will now be the ground
wire.
Solder the live wire (the one inside the clear
plastic inner tube of the Coaxial wire) to the
live
electrode wire, making sure that the two wires
running from the electrode do not touch.
Insulate the two bare connections from each
other with electrical tape.
Grounding the tank
Grounding minimizes the 60 cycle hum), use one
insulated wire stripped on both ends. Place one end into
the water, and connect the other end to the oscilloscope
(there should be a specific ground screw on the
oscilloscope, ask your teacher for directions). *Note:
all tanks that are grounded should be connected to the
same distinct ground. More than one ground will create a
ground loop.
Restraining the Fish (Follow picture diagrams.)
Make sure that the center of the fish is
directly above the center point marked on the
plexiglass grid and in line with the anchor
point.
Let fish settle for ten to twenty minutes to
ensure that the data is not altered as the fish
adjusts to its surroundings.
Use a string to attach pantyhose restraint to
the holding apparatus (see diagrams).
Procedure:
Turn on all necessary electronic equipment (oscilloscope,
amplifier, and recording equipment)
Mapping Fish Alone (control)
String setup
Mark the x-coordinate by attaching one end of
the string to the x-value on the far edge of the
tank and the other end to
the same value on the near end of the tank.
Mark the y-coordinate by attaching one end of
the string to the y-value on the left tape and the
other end to the same
value on the right tape.
Measurements
Lower the electrode at the point at which the
two strings intersect. Make sure the live wire of
the electrode is
directly below the intersection point at all
times.
Measurements when neither x nor y = 0
Place the electrode parallel to the x
axis so that the ground wire of the electrode is
farther away from the
center point than the live wire.
Place the electrode parallel to the
y-axis so that the ground wire of the
electrode is farther away from the
center point than the live wire.
Measurements when y=0 ( it is necessary to
take three vector
measurements)
x-vector: take one measurement so that
the ground wire
is along the x-axis further from the fish than the
live wire.
y-vector: take two measurements, both
perpendicular to
the x-axis, one with the ground wire
closer to the far end of the tank, and one with the
ground wore closer to the near end of the
tank.
Measurements when x = 0
x-vector: Take two measurements, both
parallel to the
x-axis, one with the ground wire closer to
the right end of the tank, and one with the
ground wire closer to the left end of the
tank.
y-vector: Take one measurement so that
the ground wire
is along the y-axis further from the fish
than the live wire.
Move the strings to the next x,y
coordinates and
repeat. repeat for every combination of x
and y coordinates as marked off by the plexiglass.
Non-conductive Object
Either suspend the glass rod from the top of the tank
or attach it to a
weight at the bottom so it extends from the bottom of the
tank to next to the fish. The
rod should be near the fish so that a distortion will be
detectable.
Repeat mapping fish alone step.
Conductive Object
Follow same procedure as for non-conductive object,
using metal
rod instead of glass rod.
Recording Measurement
On the oscilloscope, adjust the volts/cm knob and the
time/div
knob so that the entire amplitude fits on the screen.
Adjust the distance from the top tip to the bottom tip
in
centimeters (the amplitude). We found a compass works
well for this.
Record the coordinate and its amplitude along with its
amplitude along with a setting along with the setting on
the volt/cm knob.
Analysis Of Results
Find the x and y vector components for each coordinate
Multiply the amplitude by the setting of volts/ cm.
Find the magnitude of the resultant vector
When the coordinate is not x= 0 or y= 0
Take the square root of the sum of the squares
of the x and y vector components.
When x= 0
Using vector addition (paying careful
attention to signs
of vectors) add y1 and y2.
Use this vector as the y vector
component and take the square root of the sum of
the squares of the x and y vector
components.
When y= 0
Using vector addition (again paying attention
to signs) add
x1 and x2. Use this as the
x- vector component and take the
square root of the sum of the squares of the x and
y vector components.
Find the angle of the resultant vector
Take the inverse tangent of the x vector component
divided by the
y vector component.
*Note: it can be very useful and time saving to use a
spreadsheet to make
these calculations. Take the inverse tangent of the x
vector component divided by the
y vector component.
Contour graph
To see the results most clearly, graph everything on a
contour graph, plot each r value on its x and y position
coordinates.
Connect all like r values at specifics intervals
(perhaps .2 or .5)
Visually comparing the different contour graphs
The effects of the conductive and nonconductive
objects should clearly illustrated.
Possible Variations: After mapping the field at both standard and
altered levels of ion concentration, we will attempt to find whether the
field expands or collapses around the fish proportionally to ion
concentration.